Procedure: heat-flux tune

Audience: operators arming a heat-flux tune; UI authors wiring the dock; anyone debugging a stalled session from the live tick payload or audit events. Scope: the procedure surface — config fields, dock walk-through, operator commands, event taxonomy, and the live-numerics tick schema. The algorithm (convergence rule, predicate, secant step, why the constants sit where they do) is documented in Heat-flux tune procedure — the algorithm; read that page if you need to tune predicate thresholds or interpret a runaway abort.


Why this procedure exists

CAPA's scientific control parameter is the radiant heat flux at the specimen surface (kW/m²). The heater is driven by a Watlow PM3 PID closing on a temperature setpoint — temperature is a proxy for flux. This procedure runs a slow supervisory loop that discovers, for each target flux, the heater setpoint that delivers it on this rig, today.

The output is a tune artifact — a TOML file under configs/calibrations/flux/ listing (target_kw_m2, heater_setpoint_c, measured_flux) triples. Subsequent specimen runs reference this artifact via HeaterProgram.flux_calibration_ref so the heater setpoint they commanded is traceable back to the day's measured flux.

The Schmidt-Boelter gauge is removed before the specimen run. The tune is a calibration receipt, not a control law.


Quick reference

Plugin id capa.builtin.heat_flux_tune
Module src/capa/experiment/procedures/builtin/heat_flux_tune/
uses_method False — the Method tab is hidden when this procedure is selected
Required channels heat_flux_gauge, heater.setpoint, heater.pv (rebindable in config)
Bundle shape One bundle; one operational tune artifact in configs/calibrations/flux/ when persist_dir is set
Cameras / extra channels Suppressed by plan_capture — the bundle records only the three required channels

Picking config values

The HeatFluxTuneConfig schema lives in config.py with extra="forbid" — typos in field names fail validation, they do not silently default.

The fields are grouped (by capa_group metadata on each Pydantic field) so the auto-generated Run-tab editor renders them in collapsible sections. The groups, in order:

Top-level — targets and safety

Field Unit Default Notes
targets_kw_m2 kW/m² One or more target fluxes. Typically ascending ((25, 50, 75)) so the heater warms monotonically across the session. Multiple targets become a calibration sweep; single target is the daily-tune case.
t_set_max_c °C 900 Per-session hard ceiling on commanded heater setpoint. Preflight refuses anything above 1000 °C (the rig-survival limit).
operator_id None Recorded into the artifact for audit.
tolerance_kw_m2 kW/m² None Convergence band. Leave unset for the default max(0.1, 0.005 × target) per target.

Initial setpoint (initial_setpoint)

How to pick the first heater setpoint for each target. Detail in Initial setpoint heuristics.

Field Default Notes
initial_guess lookup lookup interpolates the most recent on-disk artifact. operator uses operator_initial_setpoint_c. sigma_t4 is the σT⁴ cold-start fallback.
operator_initial_setpoint_c None Required when initial_guess = "operator"; otherwise ignored. Config validation refuses the missing-pair case.

Channels (channels, advanced)

Field Default Notes
flux_channel heat_flux_gauge Name of the calibrated heat-flux gauge channel in the registry.
heater_setpoint_channel heater.setpoint Channel to issue setpoint writes against.
heater_pv_channel heater.pv Channel of the heater process variable.

The procedure reads these names — not literal strings — when filtering the recording plan, subscribing to channels, and issuing setpoint writes. Rebinding here is enough; the rest of the procedure follows.

Predicate (predicate, advanced)

The three-condition steady-state gate. Do not casually edit these. Each value sits where it does for an empirical reason — see the algorithm page before changing.

Field Unit Default What it controls
t_stable_s s 90 How long every steady-state condition must hold continuously before a measurement window is taken.
t_window_s s 180 Rolling-statistics window for mean / std / slope of flux. Sized to average across 3–4 Watlow limit cycles.
delta_t_band_c °C 0.3 Heater PV deadband: |mean(PV) − setpoint| ≤ this. Compares windowed mean, not instantaneous samples.
sigma_flux_floor_kw_m2 kW/m² 0.05 Absolute floor on the variance cap. Protects low-flux targets from chasing a cap below gauge noise.
sigma_flux_max_fraction 0.005 Relative cap on rolling std-dev as a fraction of target. 0.005 = 0.5%.
predicate_relax_factor 3.0 Distance-based relaxation. Looser slope cap + shorter dwell when far from target; tightens to 1.0 as |err| → 2 × tolerance. Set to 1.0 to disable.
slope_max_kw_per_min kW/m²/min 0.15 Maximum |d(flux)/dt|. ~20σ above the noise floor at the default window.
hampel_k 3.0 Outlier threshold in MADs over the rolling window. Hampel's classical recommendation.

Timing budgets (timing, advanced)

Interrelated — change one and the others usually need to follow. The relationships are documented in the algorithm page.

Field Unit Default What it bounds
t_settle_max_s s 1200 Per-iteration cap on time in _wait_steady. On timeout, warn and proceed with the noisier reading — does not abort.
t_verify_s s 300 Verification soak after the two in-tolerance windows. Predicate must keep holding for this dwell.
t_total_max_s s 8100 Whole-session wall-clock budget. Exhausting this aborts the current target and seals what's been accepted.
gauge_silence_max_s s 30 Abort if no fresh flux sample arrives within this window. Catches wiring/adapter failures mid-run.
poll_interval_s s 0.5 Predicate-loop wake cadence. 2 Hz is plenty since channel sample rates are 1–10 Hz.

Correction step (correction, advanced)

Defaults to a damped secant with a ±25 °C clamp. The n_iter_max default of 14 is sized for the cold-start case; warm-start sessions typically converge in 5–7.

Field Unit Default Notes
damping 0.7 Damping on the secant step. 1.0 is undamped Newton; 0.7 prevents oscillation when the prior slope is stale.
delta_t_step_max_c °C 25 Anti-runaway / human-comprehensible per-iteration ΔT clamp.
n_iter_max 14 Iteration cap per target. Exhausting it accepts the last reading with accept_reason="warn_proceeded".
runaway_sign_disagreement_count 3 Abort if sign(err) and sign(ΔT_last) disagree for this many consecutive iterations.

Safety (safety, advanced)

Field Unit Default Notes
t_safe_c °C 25 Cooldown setpoint on abort or completion. Must be below t_set_max_c; config validation enforces this.
hold_at_completion bool False Leave heater at the converged setpoint after a successful single-target tune. Four gates must all pass — see hold-at-completion. Refused at config validation for multi-target sessions.
f_gauge_sanity_max_kw_m2 kW/m² 150 Pre-loop gauge sanity ceiling. Catches calibration-off-by-10× or runaway gauge. Not a "must be cold" check — starting with the heater at an intermediate setpoint is supported.

Artifact metadata (artifact, advanced)

Field Default Notes
persist_dir configs/calibrations/flux Directory to write the artifact into. Set to blank/None to skip artifact-file persistence for this tune session.
geometry 40 mm below heater, centerline Free-text gauge-placement description recorded in the artifact.
gauge_calibration_ref None Link to the gauge's V→kW/m² calibration cert. A later analyst can detect a calibration change by comparing this across artifacts.
artifact_id_prefix capa_flux Artifact id is <prefix>_<YYYY-MM-DD> — one artifact per day per rig.

Example config

procedure:
  id: capa.builtin.heat_flux_tune
  config:
    targets_kw_m2: [25, 50, 75]
    t_set_max_c: 900
    operator_id: gbellamy
    gauge_calibration_ref: "schmidt_boelter_SN1234_cert_2026-04"
    # everything else takes defaults

A single-target daily tune is just:

procedure:
  id: capa.builtin.heat_flux_tune
  config:
    targets_kw_m2: [50]

The dock

The HeatFluxTuneDock auto-shows when RunController.active_procedure_id == capa.builtin.heat_flux_tune and hides on run completion. It surfaces three operator commands and a live-numerics panel.

The three operator commands

Button Procedure effect Audit event
Pause / Resume Sets _paused on the procedure. Settle/verify loops sleep without re-evaluating the predicate; rolling windows keep filling; heater stays at its current commanded setpoint; t_settle_max_s clock freezes. Resuming resets the predicate dwell. heat_flux_tune.operator_command with command="pause" / "resume"
Accept Current One-shot. Next poll, the loop short-circuits with the rolling-window statistics as-is. The verify soak is skipped. Resulting tune point gets accept_reason="operator_override". heat_flux_tune.operator_command with command="accept_current"
Abort Sets ctx.external_stop. The procedure raises HeatFluxTuneError from the current poll loop, then the _safe_cool path drives the heater to t_safe_c before returning. Standard run-abort audit chain

Buttons are enabled only during PREPARING, RUNNING, and DRAINING UI states — outside of those the procedure is not consuming the operator-command stream, so a click would silently no-op.

Live-numerics panel

The panel is driven by ProcedureTick payloads emitted once per poll cycle (~2 Hz at default). Numbers go visually stale (dimmed) when no tick arrives within STALE_AFTER_MS (1500 ms — two poll cycles) — catches a procedure that has stopped publishing without the dock needing to know why.

Layout, top to bottom:

Row Format What it shows
Target Target N of M · 50.00 ±0.25 kW/m² Current target, tolerance band.
Iteration Iteration 3 of 14 · SP 685.40 °C Iteration count and current commanded setpoint.
Flux 49.83 kW/m² · err −0.17 ✓ Windowed mean flux + signed error. Green check when |err| ≤ tolerance; amber otherwise.
Std 0.143 kW/m² · ≤ 0.250 Rolling std vs. per-iteration σ cap.
Slope +0.041 kW/m²/min · ≤ 0.150 Rolling slope vs. cap.
PV 685.34 °C · SP 685.40 °C Latest heater PV + commanded setpoint.
Settle dwell 78 s · 245/1200 s Predicate dwell + elapsed-of-budget.
Window 812 samples · span 175.4 s · age 0.2 s Rolling window depth, span, and last-sample age.
Phase settle · holding Current phase + predicate's last-reason string. Green when phase="holding".
In-tol 1 of 2 · runaway 0 Two-in-a-row counter; amber runaway N when the runaway detector has accumulated disagreements.
dF/dT 0.1842 · secant Jacobian used this iteration + source (secant, prior, or sigma_t4).

The predicate_last_reason string takes one of: window-not-full, missing-pv-or-setpoint, pv-out-of-band, flux-noisy, flux-drifting, holding, reset. Reading this in real time is the fastest way to know why the tune isn't accepting.


Tick payload schema

For anyone debugging the bridge directly, the ProcedureTick.payload shape this procedure publishes:

Key Type Notes
phase "settle" \| "verify_soak" \| "paused" \| "holding" Loop phase. paused is set when _paused=True regardless of which loop is running.
target_index int 1-based index within the session.
target_count int Total targets.
target_kw_m2 float Current target.
tolerance_kw_m2 float Per-target tolerance.
iteration int 1-based iteration within the current target. 0 before the first measurement.
iteration_max int n_iter_max.
commanded_setpoint_c float Latest commanded heater setpoint.
pv_latest_c float | None Most recent heater PV sample.
pv_mean_c float | None Windowed mean PV (the band gate's actual input).
mean_flux_kw_m2 float Hampel-filtered window mean.
std_flux_kw_m2 float Hampel-filtered window std.
slope_flux_kw_m2_per_min float Hampel-filtered least-squares slope.
window_full bool True once the window has accumulated ≥ t_window_s of samples.
flux_window_span_s float Wall-time span of samples currently in the window.
flux_samples_in_window int Sample count.
flux_last_sample_age_s float | None Seconds since the most recent flux sample.
predicate_dwell_s float How long the predicate has been holding (resets on any condition failing).
predicate_last_reason str See table above.
elapsed_s float Time in the current _wait_steady / _verification_soak call, paused time excluded.
settle_budget_s float t_settle_max_s during settle, t_verify_s × 2 during verify.
sigma_max_kw_m2 float The per-iteration σ cap (after distance-based relaxation).
slope_max_kw_m2_per_min float The per-iteration slope cap.
paused bool _paused flag.
in_tol_windows int Two-in-a-row counter.
df_dt_used float The dF/dT this iteration trusted.
df_dt_source "secant" \| "prior" \| "sigma_t4" Where the dF/dT came from.
runaway_count int Current runaway-disagreement count.
error_kw_m2 float | None target − mean_flux. None before the first iteration.

The holding tick (emitted once when hold-at-completion fires) carries a narrower payload: phase, target_index, target_count, target_kw_m2, commanded_setpoint_c, mean_flux_kw_m2, accept_reason, paused=False.


Audit event taxonomy

Every operator-visible state change writes a bundle_writer.write_event so the bundle records the complete session history.

Event kind When emitted Metadata of interest
heat_flux_tune.started Once, at run() entry targets_kw_m2, t_set_max_c, initial_guess
heat_flux_tune.command.issued Every setpoint write channel, device, value, accepted, authorization_id
heat_flux_tune.iteration Once per iteration iteration, target_kw_m2, setpoint_old_c, setpoint_new_c, mean_flux_kw_m2, std_flux_kw_m2, slope_kw_m2_per_min, error_kw_m2, dwell_s, dF_dT_used, dF_dT_source, decision, timed_out
heat_flux_tune.operator_command Every Pause/Resume/Accept command, operator_metadata
heat_flux_tune.target_accepted Per target target_kw_m2, setpoint_c, measured_flux_kw_m2, accept_reason
heat_flux_tune.holding When hold-at-completion fires held_setpoint_c, held_target_kw_m2, held_measured_flux_kw_m2, accept_reason, gauge_calibration_ref
heat_flux_tune.aborted On HeatFluxTuneError or wall-clock exhaustion reason
heat_flux_tune.completed Once, in the finally block accepted_points, targets_kw_m2, held

The decision field on heat_flux_tune.iteration events takes one of: step, converged_window, operator_override, abort:runaway. Reading the sequence of decision values from a bundle is the fastest way to retrace a stalled tune.


Promoting an artifact

The artifact written by this procedure is not a channel calibration. It does not feed back into a CalibrationSet; it does not auto-edit channel transforms. It is a separate file under configs/calibrations/flux/ (see Tune artifacts) that downstream consumers reference by id.

Two consumers exist today:

  1. The next heat-flux tune session. With initial_guess=lookup (the default), the procedure reads latest.toml and interpolates the previous artifact's accepted points to pick its starting setpoint.
  2. A specimen run that fills out HeaterProgram.flux_calibration_ref. This is a free-form id pointer — the operator typing capa_flux_2026-05-24 (or a notebook entry, or a calibration cert id) into the CAPA profile metadata. The pointer is recorded into the bundle for audit; capa does not auto-apply it. See CAPA profile fields.

There is no "promote artifact to calibration set" action because the two subsystems aren't transforms of the same thing. Calibration overview lays out the split.