Audience: calibration authors writing configs/calibrations/*.toml files; operators applying a set in the Setup tab; analysts parsing calibration.json out of a bundle. Scope: the on-disk TOML format of a CalibrationSet, every supported transform kind, the UncertaintySpec discipline, and the diff workflow used to apply a set to a draft.

This page covers channel-level calibration — the transform that turns a raw adapter sample into an engineering-unit channel sample. The orthogonal heat-flux tune artifact (which records a heater-setpoint↔delivered-flux mapping) is documented under Tune artifacts. See Calibration overview for why they are separate.


On-disk layout

configs/calibrations/
    sim_default.toml                  ← a CalibrationSet (this page)
    thermocouples_2026Q2.toml         ← another CalibrationSet
    flux/
        capa_flux_2026-05-24.toml     ← a HeatFluxTuneArtifact (different subsystem)
        latest.toml

One TOML file per set. Filenames are free-form; the name inside the file is what gets recorded into the bundle. Conventional naming: <channel-group>_<period>.toml (e.g. thermocouples_2026Q2.toml) or <rig>_<purpose>.toml (e.g. sim_default.toml).


Schema

A set has three top-level fields plus a [curves.*] table per channel name. The model lives in src/capa/channels/calibration.py:

name = "thermocouples_2026Q2"
revision = "3"

[curves.sample_tc_1]
kind = "linear_two_point"
input_unit = "V"
output_unit = "degC"
ref_low_raw = 0.0
ref_low_value = 0.0
ref_high_raw = 0.01
ref_high_value = 250.0

[curves.sample_tc_1.uncertainty]
kind = "absolute"
value = 0.5
coverage_factor = 2.0
method = "type-K reference + ice-point cross-check, k=2"

[curves.sample_tc_1.fit_metadata]
reference_instrument = "Fluke 1551 Stik"
reference_serial = "98765"
fitted_at = 2026-04-02T10:30:00+00:00
rms_residual = 0.18
notes = "two-point at ice and 250 °C oil bath"
Field Notes
name The set's identifier. Recorded into calibration.json so the bundle can name which set it captured.
revision Free-form revision string. Bump when curves change so a later analyst can trace which revision a bundle used.
[curves.<channel-name>] One table per channel. The key is the channel name as it appears in the hardware profile.

Every curve table carries:

Field Notes
kind Discriminator — one of identity, linear_two_point, polynomial, lookup, piecewise, custom_callable.
input_unit Wire unit. Validated against the unit registry.
output_unit Engineering unit. Validated against the unit registry; must agree with the channel's declared engineering unit at apply time.
uncertainty Optional sub-table. See Uncertainty below.
fit_metadata Optional sub-table. See Fit metadata below.
Per-kind fields See each kind below.

The dimensional algebra is checked at construction. An Identity whose units aren't dimensionally compatible, or a LinearTwoPoint whose ref_high_raw == ref_low_raw, fails validation.


Transform kinds

identity

[curves.heater_pv]
kind = "identity"
input_unit = "degC"
output_unit = "degC"

Pass-through. evaluate(raw) → raw. Required for channels whose adapter already delivers values in engineering units (Watlow temperature channels, MFC channels reporting calibrated sccm) — the data still needs a calibration entry so the bundle records "this channel was identity-calibrated" rather than "this channel is uncalibrated."

input_unit and output_unit must be dimensionally compatible. They don't need to be the literal same string (e.g. degCK could work in principle), but the runtime currently treats them as the same. Use the same unit on both sides unless you have a deliberate reason.

Identity is invertible — invert(value) → value — which matters on the setpoint write path so a derived-unit value round-trips through the identity calibration unchanged.

linear_two_point

[curves.sample_tc_1]
kind = "linear_two_point"
input_unit = "V"
output_unit = "degC"
ref_low_raw = 0.0
ref_low_value = 0.0
ref_high_raw = 0.01
ref_high_value = 250.0

Two-point linear fit: value = slope · raw + intercept. Slope and intercept are computed at construction from the two reference pairs, so the manifest and the runtime path agree (the bundle serializes ref_low_* / ref_high_*, not the derived slope).

Also invertible — used on the setpoint write path to convert a user-facing (output_unit) value back into the wire-unit (input_unit) the device expects.

polynomial

[curves.exhaust_temp]
kind = "polynomial"
input_unit = "V"
output_unit = "degC"
coefficients = [0.0, 24987.5, -0.4173]   # y = c0 + c1*raw + c2*raw²

Horner-evaluated y = c0 + c1·raw + c2·raw² + …. Coefficients are in ascending order. At least one coefficient is required (a zero-degree polynomial is a constant — degenerate but legal).

Polynomial uncertainty propagation is currently the same as the base UncertaintySpec (no Monte-Carlo per-evaluation propagation in the shipped code).

lookup

[curves.k_type]
kind = "lookup"
input_unit = "V"
output_unit = "degC"
table = [
    [0.000, 0.0],
    [0.002, 50.0],
    [0.004, 100.0],
    [0.020, 500.0],
    [0.040, 1000.0],
]

Linear interpolation over a sorted (raw, value) table. Out-of-range raws clamp to the nearest endpoint — the variant declares this explicitly rather than silently extrapolating. Extrapolated calibrations are a recurring source of "the data looked plausible" bugs; the explicit clamp produces a wrong but obvious value (the table endpoint), and the bundle's audit trail makes the clamp visible.

Validation: - The table must have ≥ 2 rows. - raw values must be strictly ascending — sorting alone is not enough; duplicate raws fail validation.

piecewise

[curves.broad_range_sensor]
kind = "piecewise"
input_unit = "V"
output_unit = "kPa"

[[curves.broad_range_sensor.segments]]
raw_min = 0.0
raw_max = 0.005
coefficients = [0.0, 200000.0]            # 0 → 1 kPa per millivolt

[[curves.broad_range_sensor.segments]]
raw_min = 0.005
raw_max = 0.020
coefficients = [-3000.0, 800000.0]        # different slope above 5 mV

Sequence of polynomial segments. Two continuity invariants are checked at construction:

  1. Adjacent segments must share their boundary raw value (prev.raw_max == next.raw_min).
  2. Adjacent segments must produce the same value at that boundary (prev.evaluate(boundary) ≈ next.evaluate(boundary) within 1e-9 × max(1, |value|)).

Discontinuous piecewise fits silently produce step artifacts in plots; the construction-time check refuses them. If you legitimately want a discontinuity (a sensor whose response changes character at a known threshold), build two separate calibrations and switch channels rather than fighting the continuity check.

Out-of-range clamps at the endpoints like lookup.

custom_callable

[curves.special_tc]
kind = "custom_callable"
input_unit = "V"
output_unit = "degC"
entry_point = "labkit.thermocouples:k_type_v3"
package = "labkit"
version = "2.4.1"
distribution_hash = "sha256:abcd..."
callable_id = "thermocouple.k_type_v3"
parameters = { ref_junction_c = 22.0 }
test_vectors = [[0.001, 25.0], [0.020, 491.5]]

Reference to an installed callable. A custom calibration must name an entry point, a package name + version, the distribution hash, a stable callable id, serialized parameters, input/output dimensions, and test vectors. Anonymous lambdas and unversioned scripts are a config error.

Field Notes
entry_point "package.module:callable" — resolvable via importlib.metadata. The : is required; missing it fails validation.
package Distribution name. Cross-checked against plugins.lock.
version PEP 440.
distribution_hash SHA-256 of the installed wheel/sdist, matched against the lockfile. Algorithm prefix (sha256: or sha512:) is enforced.
callable_id Stable id within the package (e.g. "thermocouple.k_type_v3"). The package can refactor its module layout as long as the id stays the same.
parameters Serialized parameters passed to the callable. Scalars only (float, int, str, bool).
test_vectors (raw, expected_value) pairs. The callable must reproduce these at calibration-load time — used as a self-test.

evaluate() raises if called without the plugin runtime resolved — CustomCallable validates its schema in isolation, but actually computing values requires the procedure plugin runtime. Today that callable metadata stays in the source calibration TOML and any resolved config snapshot that carries it; it is not yet copied into calibration.json.

This is the only calibration variant gated by the plugins lockfile. Built-in variants (identity, linear_two_point, etc.) don't need lockfile admission because their algebra is part of capa itself.


Uncertainty

Every calibration declares an uncertainty block — or an explicit absence. capa refuses to let an uncertainty be silently zero.

[curves.sample_tc_1.uncertainty]
kind = "absolute"           # or "relative"
value = 0.5                 # output units for "absolute", dimensionless fraction for "relative"
coverage_factor = 2.0       # k-factor: 1 = standard, 2 = ~95% expanded
method = "type-K + ice-point cross-check"
Field Notes
kind "absolute"value ± uncertainty in output units. "relative"value × (1 ± uncertainty) as a dimensionless fraction (0.01 = 1%).
value Magnitude. ≥ 0.
coverage_factor k-factor. Default 1.0. > 0. Recorded so an analyst five years later quotes the right confidence interval without guessing.
method Free-text description of how the uncertainty was estimated (residuals from a fit, manufacturer spec, …). Optional but strongly recommended.

Calibrations without an [uncertainty] block load with uncertainty = None, which is the explicit "unmeasured" state. The bundle records the None, and downstream evaluate_with_uncertainty() returns (value, None). An analyst can distinguish "we measured the uncertainty as zero" from "no uncertainty was characterised" — silent-zero is impossible.

For relative uncertainties, absolute_for(value) returns relative_fraction × |value| × coverage_factor — the dimensioned magnitude at a specific evaluation point.


Fit metadata

The pedigree of a calibration produced by a procedure (a lab calibration run, not a hand-written file).

[curves.sample_tc_1.fit_metadata]
reference_instrument = "Fluke 1551 Stik"
reference_serial = "98765"
fitted_at = 2026-04-02T10:30:00+00:00
rms_residual = 0.18
source_procedure_id = "lab.tc_two_point_fit"
capa_git_sha = "abc123..."
notes = "two-point at ice and 250 °C oil bath"
Field Notes
reference_instrument Required when this block is present.
reference_serial Optional.
fitted_at UTC datetime. Required.
rms_residual Fit residual in output units. Optional.
source_procedure_id Capa procedure id that produced this fit. Optional — hand-built calibrations don't have one.
capa_git_sha capa source commit at fit time. Optional.
notes Free-text.

The whole block is optional. Hand-typed calibrations skip it entirely. The point of having it is that a curve's origin should be recoverable without trusting human-written notesreference_instrument, reference_serial, fitted_at together name the artifact that should exist in the lab notebook.


How an experiment references a set

In the experiment YAML:

experiment:
  calibration_set: configs/calibrations/thermocouples_2026Q2.toml

At run-arm:

  1. The file is loaded via load_calibration_set.
  2. Each [curves.<name>] entry is applied to the channel of that name in the active hardware profile, overwriting whatever curve the channel originally declared.
  3. The current bundle writer records the selected set's name and revision in calibration.json. Full merged-curve snapshots are planned, but are not wired into the storage path yet.

Step 2 is destructive at apply time but non-destructive on disk — the hardware TOML stays as written; the calibration set's curves win at runtime only. This means the same hardware profile can be used with different calibration sets without editing the hardware file.


The Setup-tab apply-set diff dialog

The Setup tab's Calibration section has an "Apply set…" action that walks an operator through which curves would change before committing. The mechanics live in calibration_set_io.py.

Every (channel, set entry) pair lands in one of five classes:

DiffKind What it means Pre-checked?
override_identity Draft channel currently has an identity curve; the set provides a real curve. yes — applying cannot make things worse, the channel had no meaningful calibration before.
override_existing Draft channel has a non-identity curve; the set's curve differs. no — this is destructive (operators have lost characterisation work this way). The operator must explicitly opt in.
matches Draft channel's curve already equals the set's entry. n/a (informational, no apply checkbox).
set_only Set has a curve for a channel the draft doesn't define. n/a — there is no channel to attach it to.
channel_only Draft has a channel the set doesn't cover. n/a — the set doesn't change this channel.

The dialog renders all five classes so the operator sees the full picture, not just the rows that would change. The default-pre-checked policy (only override_identity) is deliberately conservative: a destructive overwrite of a real curve always requires a manual opt-in, every time.


In the bundle

At run-seal time the current bundle writer stores a reference snapshot in calibration.json:

{
  "name": "thermocouples_2026Q2",
  "revision": "3"
}

It does not yet serialize per-channel curves, uncertainty, or fit_metadata into the bundle. That full CalibrationSet payload is the planned shape once the calibration runtime is wired into storage.

The reference lives in calibration.json, not in manifest.json. Two bundles with the same (name, revision) are claiming to have used identical curves; bumping the revision field when you edit a set is the discipline that makes that claim trustworthy.


See also