Heat-flux tune procedure — the algorithm

Audience: anyone tuning predicate thresholds, debugging an unstable tune, or reading a stalled session in the bundle audit trail. The page that comes alongside this one is Procedure: heat-flux tune, which covers the operator/API surface — config fields, the dock, the tick payload. This page is the algorithm: the convergence rule, the predicate, the secant step, and the rationale for the empirical constants.

Scope: how capa.builtin.heat_flux_tune converges; why the tunables sit where they do; the failure modes the procedure explicitly catches.


The supervisory framing

CAPA's scientific control parameter is the radiant heat flux at the specimen surface (kW/m²). The heater itself is driven by a Watlow PM3 PID that closes on the average of three embedded thermocouples — i.e. the controller already running on the rig is a temperature loop. Heater temperature is a proxy; flux is the truth.

This procedure runs a slow supervisory outer loop on top of the Watlow's temperature PID. It does not replace the Watlow loop, it does not stream rapid setpoints, and it does not implement a flux-PID. The structure is deliberately simple:

  1. Command a heater setpoint.
  2. Wait for measured steady state.
  3. Read the windowed mean flux.
  4. Compute the error against target; apply one damped secant step.
  5. Repeat until the error sits in tolerance for two consecutive windows plus a verification soak.

The Schmidt-Boelter gauge under the heater is removed before the specimen run. The tune produces a setpoint↔flux mapping; the experiment-time controller stays temperature-only. The tune is a calibration receipt, not a control law.

The state machine lives in src/capa/experiment/procedures/builtin/heat_flux_tune/controller.py. All pure-math primitives (rolling window, Hampel rejection, the predicate, secant, runaway detector) live in the sibling signals.py so they can be unit-tested without async, a databus, or a clock.


Required hardware and channels

Three channels must be mapped on the active hardware profile:

Channel role Default name What it is
Heat-flux gauge heat_flux_gauge Calibrated kW/m² reading — typically a Schmidt-Boelter gauge wired through an NI-DAQ AI channel with the V→kW/m² curve already applied.
Heater setpoint heater.setpoint Commanded °C — the write target.
Heater PV heater.pv Live process-variable °C reading from the Watlow's TC average.

Preflight refuses to arm if any of the three is missing from the registry (see HeatFluxTune.preflight). The channel names are configurable — an operator who rebinds flux_channel = "flux_b" gets the rebinding everywhere in the procedure for free, including the recording-plan filter.

The procedure also refuses t_set_max_c > 1000 °C at preflight — that is the documented rig-survival limit, not a soft policy.


The convergence rule

                                ┌──────────────────┐
                  start  ──────►│ issue setpoint   │
                  (per target)  │ clear windows    │
                                └────────┬─────────┘
                                ┌──────────────────┐
                          ┌────►│   _wait_steady   │  ── timeout ──┐
                          │     └────────┬─────────┘               │
                          │              │ predicate fires         │
                          │              ▼                         │
                          │       take measurement                 │
                          │              │                         │
                          │       err = target − mean              │
                          │              │                         │
                  out of  │    ┌─────────┴──────────┐               │
                  tol     │    │ |err| ≤ tolerance  │ no            │
                  ◄───────┘    │ for ≥ 2 windows ?  │───────────────┤
                               └─────────┬──────────┘               │
                                         │ yes                      ▼
                                         ▼                  damped secant
                                ┌──────────────────┐         step + clamp
                                │ _verification_   │                │
                                │  soak (t_verify_s)│                │
                                └────────┬─────────┘                │
                          predicate      │ predicate       runaway? │
                          breaks ◄───────┤ holds                    │
                                         ▼                          │
                                  accept point                      │
                                  algorithm_converged               │
                                  iter cap or wall-clock ◄──────────┘
                                  → warn_proceeded

A target is accepted when two consecutive in-tolerance windows are followed by a verification soak of t_verify_s (default 300 s) during which the predicate continues to hold. The two-window rule plus the soak is intentionally redundant — a single fluke window passing tolerance does not accept; the soak then guards against an immediate predicate break that the iteration loop wouldn't otherwise catch.

Each accepted (target, setpoint, mean-flux) triple is appended to the on-disk artifact after every target, not at session end. A session that aborts on target 3 of 4 still leaves targets 1 and 2 on disk as a usable artifact — see Tune artifacts.

Three ways a target gets accepted

The accept_reason field on every HeatFluxTunePoint records which branch fired:

accept_reason Branch What it means for downstream readers
algorithm_converged Two in-tol windows + verify soak The standard happy path. Use freely.
operator_override Dock "Accept Current" button Operator decided the field was good enough. The predicate may not have held; the verify soak was skipped.
warn_proceeded Iteration cap or wall-clock exhausted The procedure could not converge in the budget. The last reading was recorded but accepted=False — interpolation helpers and "Apply latest tune" suggestions filter these out.

The accepted boolean and accept_reason are not redundant: accept_reason="warn_proceeded" always pairs with accepted=False, while the other two pair with accepted=True.


The steady-state predicate

SteadyStatePredicate is a three-condition gate with a hold-time confirmation. All three conditions must hold continuously for t_stable_s (default 90 s); any one failing resets the dwell clock to zero.

Condition Inequality What it catches
PV in band |mean(heater.pv) − setpoint| ≤ delta_t_band_c Heater is tracking. Default band is 0.3 °C.
Flux variance low std(flux) ≤ sigma_max Gauge field is quiet.
Flux slope flat |d(flux)/dt| ≤ slope_max_kw_per_min Field is not drifting monotonically.

A fourth precondition gates the whole thing: the rolling window must be warm — at least t_window_s worth of samples accumulated. Until then last_reason = "window-not-full" and the dwell clock never starts.

Why the PV gate compares means, not instantaneous samples

The Watlow PID tracks its setpoint very tightly on average but its instantaneous TC readings carry a few hundred mK of thermal and wire noise. At setpoints ≥ ~600 °C those instantaneous blips repeatedly cross a tight band and reset an instantaneous-sample timer even though the heater is steady by every other measure. Comparing the windowed mean smooths the blips out while a genuine tracking offset still moves the mean enough to trip the gate.

This is why delta_t_band_c is comfortable at 0.3 °C — it is a check on the mean, not on each raw sample.

Why a 180 s rolling window

t_window_s defaults to 180 s, which feels conservative until you watch the Watlow's closed-loop limit cycle. On this rig at high setpoints the cycle period is ~45 s. A rolling window that fits only one or two cycles aliases the cycle phase into its slope estimate, and the slope-flat gate fails ~55% of the time even when the long-term mean is dead on target.

180 s straddles 3–4 full cycles and averages the phase out. The least-squares slope estimator's 1-σ error from gauge noise drops to ~0.007 kW/m²/min at this window — about 20× headroom under the 0.15 kW/m²/min default cap, so the slope gate accepts genuine steadiness and rejects ≥ ~0.2 kW/m²/min drift cleanly.

Trade-off: the minimum useful iteration time is now bounded by the window-warm wait (~180 s). The n_iter_max and t_total_max_s defaults are sized to match — see Timing budgets.

Why the σ floor

The variance cap is max(sigma_flux_floor_kw_m2, sigma_flux_max_fraction × target). The floor (default 0.05 kW/m²) protects low-flux targets from chasing a cap below the gauge's intrinsic noise: the rig's Schmidt-Boelter head delivers σ ≈ 0.03 kW/m² on a steady field, so a 0.05 floor leaves ~1.7× headroom.

The fractional cap (default 0.5%) scales the cap upward at high targets without rewriting the config. At a 50 kW/m² target the cap is max(0.05, 0.25) = 0.25 kW/m², matching the default tolerance and leaving ~2× headroom over the ~0.13 kW/m² gauge noise observed during steady 50 kW/m² operation.

Hampel-filtered statistics

The mean, std, and slope all run over the Hampel-filtered subset of the current window (see hampel_mask). Any sample more than hampel_k MADs from the window median is dropped from the statistics; the consistency factor 1.4826 makes k=3 correspond to roughly 3σ for Gaussian noise.

The filtered samples stay in the window itself — they just don't pollute the statistics. A transient gauge glitch doesn't shorten the effective dwell once the next genuine sample lands.

Predicate strictness relaxation

For all but the first iteration the predicate caps are loosened based on distance from target. The predicate_strictness helper returns a multiplier k ∈ [1.0, relax_factor] applied as:

slope_max_effective = slope_max_kw_per_min * k
t_stable_effective  = t_stable_s / k

So when the prior iteration's error is ≥ 30% of target, the slope cap is relax_factor× looser and the dwell relax_factor× shorter — the procedure accepts a noisier reading faster because the next secant step is going to be a big move anyway, and waiting for a tight gauge field at this distance from target is wasted wall-clock. The multiplier decays linearly to 1.0 (full strictness) as |err| → 2 × tolerance.

The verify soak is never relaxed. That predicate is constructed from the config defaults directly, so the final acceptance gate always runs at full strictness even if every earlier iteration was loosened.

relax_factor = 1.0 disables the feature entirely — useful when debugging a tune that seems to be accepting too easily.


The correction step

The secant step is computed by secant_step:

ΔT_raw    = err / (dF/dT)
ΔT_damped = damping * ΔT_raw
ΔT_final  = clamp(ΔT_damped, ±delta_t_step_max_c)

Default damping is 0.7; default step clamp is 25 °C; default Jacobian default is 1 kW/m²/°C.

The function refuses to amplify a non-positive or numerically tiny dF/dTsecant_step returns 0.0 in that case rather than dividing by zero or moving the heater the wrong way. Wrong-sign Jacobians are caught one iteration later by the runaway detector.

Picking dF/dT each iteration

The Jacobian is picked by _estimate_df_dt in priority order:

  1. secant — once two in-session points exist for the current target, use the secant slope across the last two. This is the most accurate signal because it was measured on this rig in this session.
  2. prior — the prior artifact's local_df_dt around the current target, when available and positive.
  3. sigma_t4 — conservative default of 1 kW/m²/°C. Used only on iteration 1 of the first-ever tune on a rig.

The df_dt_source lands in every heat_flux_tune.iteration event and in every live tick — a tune that doesn't converge can be diagnosed by reading which Jacobian source the procedure trusted at each iteration.

Window clearing at iteration boundaries

The flux and PV rolling windows are cleared at the top of every iteration (see RollingWindow.clear). The setpoint just moved (or, for iteration 1, is about to); keeping pre-step data would dominate the std and slope statistics for the next window_s seconds and poison the predicate. The warmup wait is part of each iteration's settle budget, which is why t_settle_max_s defaults to 1200 s — leaving ~4× headroom over the ~270 s minimum useful settle (window + t_stable_s).


Initial setpoint heuristics

choose_initial_setpoint decides where to start. The choice is logged into the heat_flux_tune.started event and the result is clamped to [t_safe_c, t_set_max_c] regardless of which source wins.

initial_guess Source When right
lookup (default) Linearly interpolate against the most recent on-disk artifact; refuses to extrapolate. The normal day-to-day case where yesterday's tune is in configs/calibrations/flux/.
operator operator_initial_setpoint_c from config. An operator who knows a good starting setpoint (rig characterised by hand).
sigma_t4 σT⁴ blackbody approximation anchored at 50 kW/m² ≈ 650 °C. A truly cold start on a new rig — no artifact, no operator guess.

The σT⁴ fallback (sigma_t4_setpoint_c) is empirical, not physical. It solves F = k(T_h⁴ − T_∞⁴) for T_h with k fixed by the 650 °C / 50 kW/m² anchor on the CAPA cone. The result is a guess — iteration 1 is going to be wrong by tens of °C, the runaway detector might fire, and the procedure will hill-climb. The n_iter_max = 14 default is sized so a cold-start session has 4–5 iterations of hill-climbing before convergence still completes in the budget.

A lookup source that finds an artifact but cannot bracket the target falls through to operator, then to sigma_t4. The artifact never silently extrapolates.


Failure modes the procedure catches

Five conditions raise HeatFluxTuneError and trigger the abort path. The error is caught inside run() so the engine sees a clean return after cooldown — a HeatFluxTuneError does not crash the run.

External stop

ctx.external_stop fires from the operator's Stop button, the Run-tab Abort, or supervising code. Caught at the top of every poll loop; raises immediately. The accumulated artifact is preserved and the procedure's finally block commands the heater to t_safe_c.

Wall-clock exhaustion

t_total_max_s (default 8100 s = 2.25 h) is the entire session budget. Sized for n_iter_max=14 × ~580 s/iteration average at the 180 s window default. Single-target sessions almost always finish in 5–7 iterations; the headroom is for the multi-target sweep case.

When exhausted, the current iteration is interrupted and the artifact lands with whatever was accepted up to that point.

Settle timeout

t_settle_max_s (default 1200 s) is the per-iteration cap on time spent in _wait_steady. When elapsed without the predicate firing, the procedure warns and proceeds with the noisier reading — it does not abort the target. The resulting iteration's measurement is marked timed_out=True in the audit event and the iteration count advances. This is a deliberate choice: a tune that gets within tolerance of the target despite a noisy field is more useful than no data at all.

If every iteration times out, the iteration cap is the backstop — see warn_proceeded.

Gauge silence

If no fresh flux sample arrives for gauge_silence_max_s (default 30 s), the procedure aborts the target. Catches wiring failures, adapter crashes, and dead gauge channels mid-run. The check is last_flux_sample_ns vs. t_mono_ns() — purely passive on the sample stream.

Gauge sanity check (pre-loop)

Before the iteration loop starts, the procedure waits up to 5 s for one flux sample and verifies:

  • The value is finite (isfinite). NaN or inf → wiring/driver fault.
  • The value is below f_gauge_sanity_max_kw_m2 (default 150 kW/m²). The gauge's design full-scale is ~100 kW/m²; readings above 150 indicate a calibration off by 10× or a runaway gauge.

This is not a "must be cold" check. Starting the tune with the heater already at an intermediate setpoint is the supported workflow on this rig. The sanity check catches sensor failure modes, not heater state.

Runaway detector

RunawayDetector counts iterations where sign(err) and sign(ΔT_last) disagree. When the count reaches runaway_sign_disagreement_count (default 3), the procedure aborts with a HeatFluxTuneError.

This catches:

  • A sign-flipped prior artifact (linearly interpolated setpoint sits the wrong side of target).
  • A wiring fault where the gauge reads positive but the heater is moving the wrong direction.
  • A user who edited an artifact by hand and inverted a column.

Iterations with err == 0 or ΔT_last == 0 reset the counter — converged steps and already-clamped steps are not runaway candidates and shouldn't trip the detector.


Hold-at-completion

hold_at_completion = True leaves the heater at the converged setpoint instead of cooling to t_safe_c after a successful tune. Four gates all required (_should_hold):

  1. Operator opted in via config.
  2. The loop ran to completion of the final target. Every abort path (external_stop, wall-clock, HeatFluxTuneError) leaves completed_all_targets=False and falls through to cooldown.
  3. At least one accepted point exists.
  4. The last accepted point's accept_reason is not warn_proceeded. The artifact filters non-accepted points, so holding at a warn_proceeded setpoint would leave the operator with a hot heater whose value the rest of the system refuses to surface via "Apply latest tune" — a broken handoff. The procedure cools instead and asks for a re-tune.

A hold_at_completion=True config with more than one target is rejected at config-validation time. Holding at the final (typically highest) target of a calibration sweep is almost never the intent — the operator collecting (25, 50, 75, 100) kW/m² data ends with the heater at ~700 °C and almost always wants the default cool-down. Single-target sessions can hold; multi-target sessions cannot.

When the procedure holds, it emits a heat_flux_tune.holding event with the held setpoint, target, measured flux, and accept reason. It also publishes one final tick with phase="holding" so the dock latches into the HOLDING display before the run leaves the RUNNING state.


Timing budgets

The default budgets are interrelated — changing one usually means changing others. The relationships:

Constant Default Sized against
t_window_s 180 s Watlow limit-cycle period × 3–4
t_stable_s 90 s Half the window — long enough that a brief gauge dropout doesn't fire the predicate
t_settle_max_s 1200 s ~4× the minimum useful settle (t_window_s + t_stable_s = 270 s)
t_verify_s 300 s Long enough to catch a slow predicate break after the two in-tol windows
t_total_max_s 8100 s n_iter_max × ~580 s/iteration for the multi-target sweep case
n_iter_max 14 Cold-start session: 4–5 hill-climb iterations + 5–7 converge iterations + headroom

A tune that converges from a good prior typically finishes a single target in 5–7 iterations. A cold-start session that has to discover the rig from σT⁴ may use 10–13. If you find yourself reliably hitting the iteration cap on warm starts, the damping or delta_t_step_max_c defaults are probably wrong for the rig's heater-time-constant; tune those before raising n_iter_max.


What the bundle records

The procedure narrows the recording plan to only the three required channels and suppresses every camera (plan_capture in controller.py). The resulting bundle is a calibration receipt, not a science run — no PMMA video, no purge MFC, no sample TCs.

Event taxonomy

Event kind When emitted Carries
heat_flux_tune.started Once, at run() entry targets_kw_m2, t_set_max_c, initial_guess
heat_flux_tune.command.issued Every setpoint write channel, device, value, authorization_id
heat_flux_tune.iteration Once per iteration iteration number, target, setpoint old→new, mean / std / slope flux, error, dF_dT_used + dF_dT_source, dwell, decision
heat_flux_tune.operator_command Every dock button press command kind (pause / resume / accept_current), operator metadata
heat_flux_tune.target_accepted Per target target, setpoint, measured flux, accept_reason
heat_flux_tune.holding Once if hold-at-completion fires held setpoint, target, measured flux, accept reason, gauge_calibration_ref
heat_flux_tune.aborted On HeatFluxTuneError or wall-clock exhaustion reason
heat_flux_tune.completed Once, in finally accepted_points, targets_kw_m2, held

The decision field on iteration events takes one of: step, converged_window, operator_override, abort:runaway. Reading the sequence of decision values is the fastest way to retrace a stalled tune from the bundle.

Tick payload

The live-numerics dock subscribes to a separate stream of ProcedureTick payloads emitted once per poll cycle (~2 Hz at the default poll_interval_s = 0.5). The schema is documented inline on the dock side — see Procedure: heat-flux tune.


Operator commands mid-tune

Three commands arrive via ctx.operator_commands and mutate the procedure's state without restarting the iteration loop:

  • pause — sets _paused = True. The settle/verify loops sleep without re-evaluating the predicate, the rolling windows keep filling, the heater stays at its current commanded setpoint, and the elapsed-time clock for the settle budget freezes. Paused time does not count against t_settle_max_s.
  • resume — clears _paused. The predicate resets (its dwell clock restarts from zero) so the procedure doesn't accept a stale "holding" state.
  • accept_current — one-shot flag. On the next poll, the loop short-circuits with the rolling-window statistics as-is, marks the measurement operator_accepted=True, and the resulting tune point gets accept_reason="operator_override". The verify soak is skipped because the operator has explicitly opted out of the algorithmic convergence rule.

Unknown command kinds are logged and dropped — the procedure is forward-compatible with new OperatorCommandKind literals.


See also