Tuning workflow

Audience: operators running a heat-flux tune; calibration authors arming the procedure for the first time. Scope: the end-to-end operator path — pick a target, arm the procedure, watch convergence on the dock, save the artifact, cite it in a future specimen run.

This is the workflow page. The procedure's config surface lives in Procedure: heat-flux tune; the convergence algorithm in Heat-flux tune procedure — the algorithm; the artifact schema in Tune artifacts. This page assumes you've read Calibration overview and know which subsystem you're touching (the tune side, not the channel-cal side).


When to arm a tune

The recommended cadence is daily, per session. Heater coil emissivity drifts, gauge wiring changes between sessions, ambient temperature differs — a fresh artifact keeps the capa.flux_calibration_freshness preflight green and gives the next specimen run a calibrated starting setpoint.

Mandatory re-tune triggers:

Trigger Why
Heater coil replaced Coil age and emissivity reset; old artifact's setpoints are stale by tens of °C in some regions.
Heat-flux gauge replaced The wire-side calibration changed; old setpoints no longer correspond to the same delivered flux.
Gauge geometry changed (gauge moved, holder swapped, fixture redesigned) The flux-vs-setpoint curve depends on the gauge's position relative to the heater.
New rig profile A fresh hardware profile has no on-disk artifact at all.
The freshness preflight warning fires The lab-policy recency window (default 7 days) has elapsed.

Procedure choices

Today there is one tune procedure: capa.builtin.heat_flux_tune. The plan accommodates future variants (e.g. a per-band emissivity tune, a holder-geometry sweep) but none ship yet.

Choose initial_guess based on what you have on hand:

Source Use when
lookup (default) A recent artifact is on disk under configs/calibrations/flux/. The most common case — yesterday's tune is the prior.
operator You're characterising a new rig by hand and know a safe starting setpoint. Requires operator_initial_setpoint_c.
sigma_t4 Truly cold start, no artifact, no operator knowledge. The σT⁴ fallback is anchored at 50 kW/m² ≈ 650 °C and will be wrong by tens of °C — the procedure hill-climbs in iterations 1–4.

Defaults are tuned for the daily-tune workflow: 1 target ([50]), lookup priors, 900 °C ceiling. For a calibration sweep, change targets_kw_m2 to the ascending list of points you want, e.g. [25, 50, 75, 100].


Preflight checklist

Before arming, confirm:

  1. The three required channels are mapped on the active hardware profile:
  2. heat_flux_gauge (or whatever flux_channel you set)
  3. heater.setpoint (or whatever heater_setpoint_channel)
  4. heater.pv (or whatever heater_pv_channel)

Missing channels are caught by procedure preflight with code="heat_flux_tune.missing_channel".

  1. The Schmidt-Boelter gauge is physically installed at the operator-declared geometry position. The procedure's gauge-sanity check catches NaN/garbage but does not know whether the gauge is in the right place — that's an operator judgment call.

  2. t_set_max_c ≤ 1000 °C. Preflight refuses higher; even if it didn't, that's the documented rig-survival limit.

  3. Cooling water + safety interlocks are live. The tune runs the heater for tens of minutes at hundreds of °C — same physical hazards as a specimen run.

  4. No specimen is on the load cell. Run the tune with the load cell unloaded (or with a known thermal mass) but never with the day's actual specimen — the heat-up profile would consume the sample before the real run.


Arming from the GUI

  1. Open the Setup tab.
  2. Select procedure: Heat-Flux Tune. The Method tab disables itself (uses_method=False).
  3. Edit the procedure config in the auto-generated form. The form groups fields by capa_groupTop-level (targets + safety) is always expanded; advanced groups (predicate, timing, correction, safety, artifact) default to collapsed. For a daily tune, Top-level is usually all you touch.
  4. Set operator_id and gauge_calibration_ref — both land in the artifact and are read at audit time. Skipping gauge_calibration_ref makes it harder to detect a future calibration change.
  5. Click Arm. Standard preflight runs; missing channels or invalid configs surface before the run starts.
  6. Hit Run.

The HeatFluxTuneDock auto-shows when the run state transitions to RUNNING.


Arming headless

uv run capa run --headless tune.yaml

with tune.yaml:

experiment:
  hardware: configs/hardware/capa_real_full.toml
  domain_profile:
    id: capa.profiles.capa_pyrolysis  # optional but recommended
  sample:
    id: tune_2026-05-24
    notes: "Daily tune, all-new gauge cable"

procedure:
  id: capa.builtin.heat_flux_tune
  config:
    targets_kw_m2: [50]
    operator_id: gbellamy
    gauge_calibration_ref: schmidt_boelter_SN1234_cert_2026-04

The dock is not active in headless mode (no UI), so the three operator commands (Pause / Resume / Accept Current) are unavailable. The procedure runs to completion or until the wall-clock budget exhausts. Audit events + the on-disk artifact land normally.


Watching convergence on the dock

The live-numerics panel updates at ~2 Hz. The two fields that tell you the most:

  • Phase + last reason. phase shows the loop you're in (settle, verify_soak, paused, holding). The last-reason column shows the predicate's current state: holding, pv-out-of-band, flux-noisy, flux-drifting, window-not-full, missing-pv-or-setpoint. A predicate stuck on flux-noisy after the window is warm means the variance cap is tight for the current rig conditions — see Troubleshooting below.
  • Flux + err. Green check when |err| ≤ tolerance, amber otherwise. Two consecutive green ticks + the verify soak → acceptance. A flux value that oscillates around target without ever staying green for t_stable_s (default 90 s) usually means a noisy gauge field — the iteration eventually settles or times out into warn_proceeded.

The In-tol row counts windows of 2 — the two-in-a-row acceptance counter. When it reads 2 of 2, the procedure is in the verify soak.

The dF/dT row's source field (secant, prior, or sigma_t4) is the fastest way to see whether the procedure is using a good prior or hill-climbing. secant from iteration 2 onward is the happy state.


When to abort an unstable tune

Hit Abort if any of the following fire:

Symptom Likely cause Next step
predicate_last_reason = "flux-noisy" persistently after the window warms Gauge cable, EMI source nearby, or cabinet door open Check physical setup; if persistent, widen sigma_flux_floor_kw_m2 or sigma_flux_max_fraction and re-run
predicate_last_reason = "flux-drifting" with a real ramp visible in the slope row Heater not yet at steady state; needs more dwell Usually self-corrects — the predicate-relaxation logic gives the procedure a few iterations to get close
runaway_count reaching 2 (one short of the trip) Wrong-sign Jacobian, sign-flipped prior, gauge wired backwards Abort, check gauge_calibration_ref and the gauge wiring before retrying
Flux reading abruptly drops to ~0 or pegs at the gauge full-scale Gauge cable disconnected, water-cooling line clogged Abort immediately — the procedure's gauge-silence check will fire in 30 s but a clogged cooling line damages the gauge
Heater PV diverges from setpoint by > 10 °C with no Watlow error Watlow controller fault Abort; investigate via Watlow's own front panel

The procedure's own safety nets (gauge silence, gauge sanity, runaway detector, wall-clock budget) catch most of these eventually. Hitting Abort is the operator-friendly path when you see it sooner than the procedure does.

After an Abort, the procedure's finally block drives the heater to t_safe_c (default 25 °C). Any accepted targets up to the abort point are preserved in the on-disk artifact.


Where the artifact lands

One artifact file is written when persist_dir is configured:

Path Purpose
configs/calibrations/flux/<id>.toml (default persist_dir) Operational copy. Read by the next tune session's lookup prior and by the CAPA profile UI.

The id is <artifact_id_prefix>_<YYYY-MM-DD> (default prefix capa_flux, so capa_flux_2026-05-24.toml). latest.toml is repointed at the new id; the previous target is copied to <previous-id>.toml.bak-<today> first. If persist_dir is null, artifact-file persistence is skipped. The tune run's bundle records config and audit events, but it does not currently include an automatic artifact sidecar. Details in Tune artifacts.


Citing the artifact in a specimen run

After the tune, the next specimen run's CAPA profile metadata cites the artifact by id:

domain_profile:
  id: capa.profiles.capa_pyrolysis
  metadata:
    program:
      target_heat_flux_kw_m2: 50.0
      heater_setpoint_c: 727.0           # operator types this from the artifact
      flux_calibration_ref: capa_flux_2026-05-24

Two facts worth being explicit about:

  1. The reference is free-form. capa does not parse flux_calibration_ref as a file path or auto-load the artifact at run-arm. It's metadata for the bundle.
  2. heater_setpoint_c is operator-typed, not auto-derived. The Setup-tab UI offers to populate it from the latest artifact (using setpoint_for_target), but the operator confirms the value before arming. This is deliberate — an operator who knows the morning's tune was suspect can override.

There is no automatic "promote artifact to calibration set" action because the two subsystems are unrelated. See Calibration overview for the framing.


Troubleshooting

"The tune kept timing out into warn_proceeded"

Symptoms: every iteration's dwell_s is close to t_settle_max_s; last_reason stays on flux-noisy or flux-drifting.

Likely causes, in order of frequency:

  1. The gauge field is genuinely noisier than the cap. Check that the cabinet door is closed, no fans pointing at the rig, no cycling HVAC. The default sigma_flux_max_fraction = 0.5% and sigma_flux_floor_kw_m2 = 0.05 are tight on this rig; a noisier rig may need looser values.
  2. A new gauge cable was just installed without retuning the predicate. New cabling can change the noise envelope. Run a deliberate look at std_flux for ~10 minutes at a known steady setpoint (free run, manual heater command) to characterise the noise before re-tuning.
  3. t_window_s is too small for the heater's current limit-cycle period. Defaults to 180 s, sized for the rig's observed ~45 s cycle period. If the rig is closing on a longer cycle (different setpoint, different gain), widen the window.

"Iteration 1 wildly overshoots the target"

Symptoms: sigma_t4 source in iteration 1, then a runaway warning by iteration 3.

The σT⁴ fallback is anchored at one point — it can be off by tens of °C for very high or very low targets. The procedure's predicate_relax_factor is meant to absorb this (loose predicate when far from target → next iteration's secant correction is large but informed). If the runaway detector trips before convergence, the most reliable fix is to set initial_guess = "operator" and supply an operator_initial_setpoint_c close to the expected value.

A historical artifact (even from a different rig, even from months ago) is usually a better prior than σT⁴. If one exists, set initial_guess = "lookup" and point persist_dir at the artifact's directory.

"The tune converged but the held setpoint looks wrong"

Symptoms: artifact accepts at a setpoint visibly different from what you'd expect from the historical record.

Read the artifact: does gauge_calibration_ref match yesterday's? If not, the gauge's V→kW/m² curve changed under the tune — yesterday's and today's setpoints are not directly comparable.

If gauge_calibration_ref matches and the setpoint is still off by tens of °C, suspect physical changes: the gauge moved (geometry no longer the recorded 40 mm below heater, centerline), the holder was swapped, a different specimen disk was previously left in the field.

"I aborted partway through a multi-target session — do I have to start over?"

No. Any targets accepted before the abort are in the on-disk artifact. The next session reads it as the lookup prior; if you only need to re-do the failed targets, edit the session's targets_kw_m2 to that list. The partial-save behavior is documented in Tune artifacts.